The Fluke 719PRO provides an innovative, built-in electric pump and pressure calibration that is accurate and easy to use. Now you can calibrate and test pressure devices quickly and easily with one hand, saving you valuable time. The Fluke 719Pro's programmable pump limit settings to eliminate over-pressurization, and to set ‘pump to’ pressure values. The unique cleanout port design reduces pump failures, and makes the Fluke 719Pro easy to clean without repair. The compact, lightweight form, combined with one-handed use, makes this pressure calibrator the ideal tool for busy process professionals. The Fluke 719 Pro adds a backlit display and adds precision temperature measurement with the use of a RTD probe, making the Fluke 719 Pro a very veritile Calibrator.
Specifications | |||||
Pressure Measurement Specifications | |||||
Range | 719PRO-30G | -12 to 30 PSI /-0.8 to 2 bar | |||
719PRO-150G | -12 to 150 PSI /-0.8 to 10 bar | ||||
719PRO-300G | -12 to 300 PSI / -0.8 to 20 bar | ||||
Resolution | 0.01 PSI, 0.001 bar | ||||
Accuracy (1 year) | ±0.025 % of full scale, 6 months ± 0.035 % of full scale, 1 year Dry Air and non-corrosive gas only | ||||
Temperature effect (all ranges) | Add ± 0.002 % F.S./°C for temps outside of 15°C to 35°C No effect on accuracy on all functions from 15°C to 35°C |
||||
Function | Range | Resolution | Accuracy (1 year) | ||
mA dc (simulation and measurement) | 0-24 mA dc | 0.001 mA | ±0.015 % of reading ±2 counts |
1000 ohm maximum load in mA source, 26 V dc maximum voltage in mA simulate | |
Volts dc (measurement only) | 30 V dc | .001 V | ±0.015 % of reading ±2 counts |
No AC voltage measurement, do not exceed 30 V dc | |
Temperature (measurement only, Pt100 RTD) | -50 to 150°C (-50°F to 302°F) | 0.01°C 0.01°F |
Temperature ± 0.1°C (0.2°F) | ±0.25°C (±0.45°F) combined uncertainty when using 720 RTD probe (optional accessory) | |
Loop power supply | 24 V dc | N/A | 24 mA at 24 V | ||
Temperature effect (all functions) | Add ± 0.002 % F.S./°C for temps outside of 15°C to 35°C | No effect on accuracy on all functions from 15°C to 35°C | |||
Environmental Specifications | |||||
Operating temperature | -10 °C to +50 °C | ||||
Storage temperature (with batteries) | Per battery manufacturer's specification within storage specification below | ||||
Storage temperature (without batteries) | -20 °C to +60 °C | ||||
Power Requirements | |||||
Battery | Eight (8) standard Alkaline AA batteries | ||||
Battery life | 300 pump cycles to 150 PSI minimum; 1000 pump cycles to 30 PSI; 100 pump cycles to 300 PSI minimum | ||||
Physical Specifications | |||||
Dimensions (HxWxD) | 9.5 in x 4.4 in x 2.8 in (24.13 cm x 11.18 cm x 7 cm) | ||||
Weight | With holster 1.2 kg (2.5 lb) | ||||
Connections | |||||
Pressure | 1/8 inch female NPT | ||||
Electrical | Standard banana jacks | ||||
RTD | 4 pin miniature circular | ||||
External module | 6 pin miniature circular | ||||
Connections | |||||
psi, bar, mbar, kPa, kgcm2, cmH2O@4°C, cmH2O@20°C, mH2O@4°C, mH2O@20°C, inH2O@4°C, inH2O@20°C, inH2O@60°F, mmHg@0°C, inHg@0°C, ftH2O@4°C, ftH2O@20°C, ftH2O@60°F |